Endoscopic Ultrasound

ORIGINAL ARTICLE
Year
: 2019  |  Volume : 8  |  Issue : 1  |  Page : 50--57

A retrospective histological comparison of EUS-guided fine-needle biopsy using a novel franseen needle and a conventional end-cut type needle


Shuntaro Mukai1, Takao Itoi1, Hiroshi Yamaguchi2, Atsushi Sofuni1, Takayoshi Tsuchiya1, Reina Tanaka1, Ryosuke Tonozuka1, Mitsuyoshi Honjo1, Mitsuru Fujita1, Kenjiro Yamamoto1, Yukitoshi Matsunami1, Yasutsugu Asai1, Takashi Kurosawa1, Yuichi Nagakawa3 
1 Department of Gastroenterology and Hepatology, Tokyo Medical University, Tokyo, Japan
2 Division of Diagnostic Pathology, Tokyo Medical University, Tokyo, Japan
3 Third Department of Surgery, Tokyo Medical University, Tokyo, Japan

Correspondence Address:
Dr. Takao Itoi
Department of Gastroenterology and Hepatology, Tokyo Medical University, 6-7-1 Nishishinjuku, Shinjuku-Ku, Tokyo 160-0023
Japan

Background and Objectives: Recently, a 22G Franseen needle for EUS-guided fine-needle biopsy (EUS-FNB) with three novel symmetric heels has been developed to adequately obtain a core tissue. Methods: All 38 consecutive patients with pancreatic masses who underwent EUS-FNB using a Franseen needle were investigated retrospectively to assess the efficacy and safety of EUS-FNB using the Franseen needle. Then, the EUS-FNB outcomes and histological assessments of the tissue obtained by EUS-FNB using the Franseen needle and EUS-FNA using the conventional end-cut type needle for each of the 30 pancreatic ductal adenocarcinoma cases were compared. Results: An accurate histological diagnosis of the Franseen needle was achieved with a mean of 2 passes in 97.4% of patients. Although the accurate histological diagnosis rate of pancreatic ductal adenocarcinoma was not significantly different (96.7% vs. 93.3%, P = 0.55), the mean number of passes in the Franseen needle was significantly less than that in the conventional needle (2.1 ± 0.4 vs. 3.2 ± 0.8, P < 0.001). The presence of desmoplastic fibrosis with neoplastic cellular elements and venous invasion were significantly higher (96.7% vs. 40.0%, P < 0.001 and 23.3% vs. 0%, P < 0.01, respectively) and the amount of obtained tissue was significantly larger with the Franseen needle (2.13 mm2 vs. 0.45 mm2, P < 0.001). Conclusions: EUS-FNB using the Franseen needle enables the acquisition of a larger amount of tissue sample and achieves an accurate histological diagnosis with a smaller number of passes than the conventional end-cut type needle.


How to cite this article:
Mukai S, Itoi T, Yamaguchi H, Sofuni A, Tsuchiya T, Tanaka R, Tonozuka R, Honjo M, Fujita M, Yamamoto K, Matsunami Y, Asai Y, Kurosawa T, Nagakawa Y. A retrospective histological comparison of EUS-guided fine-needle biopsy using a novel franseen needle and a conventional end-cut type needle.Endosc Ultrasound 2019;8:50-57


How to cite this URL:
Mukai S, Itoi T, Yamaguchi H, Sofuni A, Tsuchiya T, Tanaka R, Tonozuka R, Honjo M, Fujita M, Yamamoto K, Matsunami Y, Asai Y, Kurosawa T, Nagakawa Y. A retrospective histological comparison of EUS-guided fine-needle biopsy using a novel franseen needle and a conventional end-cut type needle. Endosc Ultrasound [serial online] 2019 [cited 2019 Oct 20 ];8:50-57
Available from: http://www.eusjournal.com/article.asp?issn=2303-9027;year=2019;volume=8;issue=1;spage=50;epage=57;aulast=Mukai;type=0